4.3 Check weighing
For this purpose, F2.1=OVEr and F2.2.1=CHECh (factory setting) must be set in the op-
erator menu. In the factory setting, the check weighing function is working with upper and low-
er tolerances of 10 d. With parameters F2.2.3 and F2.2.4, these tolerances can be custom-
ized.
Setting target weight
1. Press
to activate the check weighing function.
2. Press and hold
If F2.2.2=WEIGHt (factory setting) is set in the operator menu, the weight display appears.
3. Put the target weight on the weighing platform and save with
The OK indicator lights.
If F2.2.2=MAnUAL is set in the operator menu, the weight display with blinking last digit
appears.
3. Enter target weight using the
page 15).
4. Save entered weight value as target weight using the
Check weighing
Example: Target weight = 1.000 kg
Switching between checkweighing and normal weighing
➜ Press
to switch between checkweighing and normal weighing.
until tArGEt and the 3 indicators Under, OK and Over appear.
,
and
keys and confirm with
key.
• Weight less than target weight and below
lower tolerance.
The Under indicator lights.
• Weight within tolerances.
The OK indicator lights.
• Weight more than target weight and
above upper tolerance.
The Over indicator lights.
12
.