Procedure for models SL415C, SL415L and SL4201C
1) Fit the universal Ivory turbine handpiece adapter to
the oil spray can. Insert the other end of the universal
Ivory adapter into the chuck in the handpiece. Hold the
handpiece in place. Spray for about 5 seconds.
2) Fit the adapter for Ivory heads and contra-angles to
the oil spray can. 3) Apply the adapter to the back of the
handpiece. Hold the handpiece in place and spray for
about 5 seconds.
CAUTION!
Incorrect or inadequate spindle cleaning can cause the
spindle to loosen and accidentally release the bur during
functioning.
CAUTION!
After lubricating, hold the handpiece with the chuck
facing down. Operate the handpiece for 30 seconds to
expel any excess oil. If you notice any dirt coming out,
repeat the entire cleaning and lubrication process.
53
Eng