Figure 32
√ Kinetics
Upper limit
Lower limit
Display data
U
TILITIES
On the main menu display select Utility with ▼/▲ (2.9) and press ENTER (2.7). A list with all the utilities of the
equipment will be displayed (Fig. 33).
Use ▼/▲ (2.9) to browse all utilities and press ENTER (2.7) to go into the selected function.
Press RETURN (2.6) to go back to the main menu.
Figure 33
Basic mode
Quantitative
Kinetics
Utility
√ Utility
Date & time
Dark current
Reset WL.
1. Deuterium lamp on/off (D2 Lamp On/Off) (Fig. 34)
Note: Zuzi recommends to have the deuterium lamp switched off when working out of the ultraviolet range.
On the function selection display, move the cursor with ▼/▲ (2.9) to select D2 Lamp and press ENTER (2.7) to
enter the deuterium lamp state selection display.
By using ▲▼ (2.9) select On or Off and press ENTER (2.7). Beside the selected state the symbol √ will be displa-
yed.
Press RETURN (2.6) to go back to the function selection display.
Note: When the deuterium lamp is switched on, it will take a few seconds to warm up.
Figure 34
√ Utility
D2 Lamp
W Lamp
Printe
2. Tungsten lamp on/off (W Lamp On/Off) (Fig. 35)
Note: Zuzi recommends to have the tungsten lamp switched off when working out of the visible range.
On the function selection display, move the cursor with ▼/▲ (2.9) to select W Lamp and press ENTER (2.7) to
enter the deuterium lamp state selection display.
By using ▲/▼ (2.9) select On or Off and press ENTER (2.7). Beside the selected state the symbol √ will be displa-
yed.
Press RETURN (2.6) to go back to the function selection display.
Page. 30
ENTER
RETURN
ENTER
15:15
04/21
RETURN
√ Utility
Lamp life
Load default
Lamp change.
ENTER
on/off
on/off
RETURN
Instruction manual 54251050
No.
Abs
1
0.000
2
0.001
3
0.002
4
0.003
√ Utility
D2 Lamp
on/off
W Lamp
on/off
Printe
√ Utility
√ D2 Lamp
ON
√
OFF
Version
Version 2 Nov-10