Error Codes
Error codes are shown on the main display. "Err" codes are shown continuously;
"Inf." messages are shown for 2 seconds,
after which the program returns automatically to the weighing mode.
Problem
Nothing appears
– No power available
or some symbols
– The AC adapter is not plugged in
incomplete
– Battery is drained
on the display
– LCD cable is loose
– The load exceeds the scale capacity
oL
uL
– Weighing pan is not in place
– Something is touching the weighing pan – Move the object that is touching the
d5PErr
– Display overflow:
CALErr
– Calibration parameter not met; e.g.:
APPErr
– Weight is too high or there is no sample – Increase the weight on the scale
PrtErr
– Data interface for printing is blocked
– APP board cannot retrieve data
5y5Err
EPErr
– EEPROM defective
FnErr
– Function not allowed in scales
5b1Err
– SBI command error
– Zero is not possible if weight value
Err 08
– Tare is not possible when the gross
Err 09
– Fatal error; cause unknown
FEErr
Err 54
Err 55
– No platform is connected
Max. Weighing capacity – The scale was switched on without
is less than indicated
under "calibrate"
The weight readout
– The scale was not calibrated/
is obviously wrong
– Scale not zeroed
Cause
Value cannot be shown on the display
– scale not zeroed
– scale is loaded
– scale setting not allowed
to the customer
on the scale with application in use
from AD board
verified for use in legal metrology
is more than 10% of FSD
weight is a minus value
the weighing pan in place
adjusted before weighing
Solution
– Check the power supply
– Plug in the AC adapter
– Replace battery; charge battery using
external charger
– Plug in the cable again or contact
customer service
– Unload the scale
– Place the weighing pan on the pan
weighing pan
– Reduce the load on the scale
– Calibrate only when zero is displayed
– Press (zero) to tare the scale
Unload the scale
– Contact customer service
– Contact customer service
– Contact customer service
– Contact customer service
– Contact customer service
– Zero the scale after removing
some of the load
– Zero the scale
– Contact customer service
– Connect a platform or sensor
– Place the weighing pan on the scale
and press "On/Off"
– Calibrate/adjust the scale
– Zero the scale
31