Seca 676 Manual De Instrucciones Y Garantía página 27

Ocultar thumbs Ver también para 676:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 58
8. What do I do if...
... no weight display comes on?
– Is the scale switched on?
– Is the plug contact of the mains unit properly connected to the scale?
– Is the plug contact for connecting the display to the scale correctly connected?
0. 0
...
does not appear before weighing?
– Press the Start key again – there must not be any load on the scale – and only its feet
should be in contact with the floor.
... one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.
...
St0p
appears in the display?
– Maximum load has been exceeded.
... the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the scale
0. 0
and wait until
... the display
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
E
... the display
and a number appear?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, disconnect the power supply by briefly removing the mains unit.
If this measure is equally unsuccessful, inform the Service department.
9. Servicing
Your seca scale leaves the factory with an accuracy greater than ±0.15 %. To ensure that
this accuracy is maintained, the product must be carefully set up and regularly maintained.
Depending on the frequency of use, we recommend servicing every 3 to 5 years. Please
ensure that a qualified service agent is used. If in doubt, contact seca Customer Service for
further information.
Model 676
is displayed, then weigh again.
appears?
27
GB

Publicidad

Tabla de contenido
loading

Tabla de contenido