Guidance and Manufacturer's Declaration - Electromagnetic Immunity:

The P.R.O. Matt Plus Control Unit (PMP-CU) is intended for use in the electromagnetic environment (for home
healthcare and professional healthcare) specified below. The customer or the user of the P.R.O. Matt Plus Control
Unit (PMP-CU) should assure that it is used in such an environment.
Immunity Test
IEC60601 Test Level
Electrostatic
Contact: ±8 kV
discharge (ESD)
Air±2 kV,±4 kV,
IEC 61000-4-2
±8 kV,±15 kV
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±0.5kV,±1kV line(s)
to line(s)
IEC 61000-4-5
±0.5kV,±1kV,±2kV
line(s) to earth
Voltage
Voltage dips:
dips, short
interruptions
0% Uτ; 0,5 cycle
and voltage
variations on
0% Uτ; 1 cycle
power supply
input lines IEC
70% Uτ; 25/30 cycles
61000-4-11
Voltage interruptions:
0% Uτ; 250/300
cycle
Power
30 A / m
frequency (50,
60 Hz) magnetic
50 Hz or 60 Hz
field IEC 61000-
4-8
Note: Uτ is the a.c. mains voltage prior to application of the test level.
*During DIP interference, the pump will outage these are normal. The pump outage does not affect the motor
operation.
Compliance Level
Electromagnetic Environment-Guidance (for
home healthcare and professional healthcare
environment)
Contact: ±8 kV
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
Air±2 kV,±4 kV,
relative humidity should be at least 30%.
±8 kV,±15 kV
±2 kV for power
Mains power quality should be that of a typical
supply lines
home healthcare and professional healthcare
environment.
Not applicable
±0.5kV,±1kV line(s)
Mains power quality should be that of a typical
to line(s)
home healthcare and professional healthcare
environment.
Not applicable
Voltage dips:
Mains power quality should be that of a typical
home healthcare and professional healthcare
0% Uτ; 0,5 cycle
environment. If the user of the P.R.O. Matt Plus
Control Unit (PMP-CU) requires continued
0% Uτ; 1 cycle
operation during power mains interruptions, it is
recommended that the P.R.O. Matt Plus Control
70% Uτ; 25/30 cycles
Unit (PMP-CU) be powered from an uninterruptible
power supply or a battery.
Voltage interruptions:
0% Uτ; 250/300
cycle
30 A/m
The P.R.O. Matt Plus power frequency magnetic
fields should be at levels characteristic of a
50 Hz, 60 Hz
typical location in a typical home healthcare and
professional healthcare environment.
Joerns® Support Surfaces
P.R.O. Matt® Plus
© 2020 Joerns Healthcare • 6110394 RevC • 20-2539
15