Powered Modes

Patient Comfort Controls and Monitoring
The P.R.O. Matt
Plus converts to an alternating
®
pressure redistribution system when connected to a

control unit.

Power Button
Power button is used to turn the power on
and off.
Standby Light
The Standby light illuminates when the unit is initially
plugged in, indicating power is available. If the unit is
turned off by pressing the Power button, the Standby
light re-illuminates. If the Standby light illuminates
without a caregiver pressing the Power button, it
indicates that there has been a power interruption and
the therapy control unit is ready to be turned back on.
Press the Power button and reset the preferred mode
of therapy and comfort level.
Autofirm
Autofirm mode provides maximum air
inflation designed to assist both patients
and caregivers during patient transfer and treatment.
Pressing the Autofirm button will illuminate the On
light. The unit will automatically return to the mode it
was in prior to Autofirm in approximately 15 minutes.
Control Unit (Figure 2)
The control unit has two therapy modes: Static and
Alternate. The Mode button is used to place the control
unit in one of the desired modes of operation.
• Static: The Static light is illuminated when in Static
mode. This mode should be used along with the
Comfort Adjust buttons to increase or decrease the
general firmness of the P.R.O. Matt Plus mattress.
• Alternate: The Alternate light is illuminated when
in the Alternate mode. The pressure automatically
alternates between adjacent cells. The alternate
frequency is carried out at preset time intervals set
by the Cycle Time button.
8
© 2020 Joerns Healthcare • 6110394 RevC • 20-2539
Cycle Time
When the unit is placed in Alternate mode,
the five-minute time light illuminates and the
mattress therapy cells will alternate in pressure on a
five-minute cycle. The cycle times can be adjusted
to 5, 10, 15, and 20-minute intervals by pressing the
Cycle Time button.
Comfort Adjust
The Comfort Adjust buttons are located on
the right of the control panel. Use the plus
and minus buttons to increase or decrease
general firmness in the mattress.
Lockout/Silence
The unit is designed to lockout all the
adjustment controls after the patient has
been positioned correctly. Approximately five minutes
after the last button is pushed, the Lockout is enabled,
and the Lock light will illuminate. This feature is to
prevent any unauthorized changes to the patient
settings. To unlock and make adjustments to the
settings, press and hold the Lockout/Silence button.
Lockout mode will return after approximately five
minutes of inactivity.
The audible warning can be temporarily muted by
pressing the Lockout/Silence button on the front panel
– audible warning will be muted for 20 minutes.
Low Pressure Indicator
The Low Pressure light will illuminate and an audible
warning will occur if there is not enough pressure
in the inner air cells. If this occurs, check the hose
connection to the mattress to ensure the hoses are
tightly connected without air leakage.