Processing; Cleaning/Disinfecting - Aesculap Suhl FavoritaCL Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 48
5.

Processing

Damage to, or destruction of the
product
cleaning/disinfecting or steriliza-
tion!
CAUTION
Only clean and disinfect the
product manually.
Do not sterilize the product
under any circumstances.

5.1 Cleaning/Disinfecting

Note
Always follow disposal and hygiene instructions and
regulations!
Manual cleaning/disinfecting
Risk of electric shock and fire
hazard!
Unplug the unit before cleaning.
DANGER
Make certain that no fluids will
penetrate the product.
Leave the product exposed to
air for at least 1 minute after
cleaning/disinfecting.
caused
by
mechanical
Damage to the product due to
incorrect cleaning or inappropriate
cleaning/disinfecting agents!
Use any commercial detergent
CAUTION
for surface cleaning.
Never immerse the product in
water or detergent.
Do not introduce any fluid into
the charging bays.
Clean
the
charging bays, applying extreme
care.
Wipe the shell of the product with a lint-free cloth
moistened with a commercial disinfectant for wipe
disinfection.
If necessary, after the appropriate exposure time,
wipe off any residues of the cleaning/disinfecting
agent with a lint-free cloth moistened with clean
water.
Dry the product with a clean, lint-free cloth.
Clean the contacts in the charging bay with
isopropanol or ethyl alcohol and a cotton swab. Do
not use corrosive chemicals.
If necessary, repeat the cleaning/disinfecting
process.
contacts
in
the
9

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Libra

Tabla de contenido