Dynamic E-CHART Manual De Operación página 138

Computador de flujo
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Dynamic Fluid Measurement
DATA PACKET
Example Modbus Previous Hour Flow Data Collection
Query - 7099 Hourly Pointer
Address
Func
01
03
ECHART Response.
Address
Func
01
03
Query - 704 - Previous Hourly Flow Data Record 1 – Last Hour Data
Address
Func
01
03
ECHART Response–Record 1:the last hour data
Address
Func
01
03
Archive
Record
704
1
E-CHART Manual
Start HI
Start LO
1b
bb
Bytes
Data Hi
04
Start HI
Start LO
02
bf
Bytes
Data
2c
46ff7e00
48609c00
423e1111
42840000
43fa0000
42096b52
449bd99a
42630674
449bd99a
42c80000
4333aaee
Description
Date
Time
Flow Time (Minutes)
Average Pressure
Average Temperature
Gross Total
Net Total
Mass Total
Energy
Average DP
Average DP/EXT
www.dynamicflowcomputers.com
No of Register
CRC
00
01
Data Lo
CRC
Record Number
CRC
00
01
CRC
fd
Value
3/27/03
23:00:00
47.51
70.0
34.555
1246.8
56.75
1246.8
100.0
179.668
Modbus Data – 4-52