Product Description

The GlideRite
®
Single-Use Stylet was designed to help enable the placement of an endotracheal tube (also known as an
ETT or ET tube). The rigidity of this stylet helps the user manipulate the tube as desired for intubation. The stylet is for
use in ET tubes between 3.0 and 4.0 mm in internal diameter.
Statement of Intended Use
The GlideRite Single-Use Stylet is an intubation-assist tool intended for use by qualified medical professionals,
for guiding an endotracheal tube past the vocal cords into the trachea, shaped to work with GlideScope
laryngoscope products.
Statement of Prescription
Caution: Federal (U.S.) law restricts this device to sale by or on the order of a physician.
WARNINGS
• Inspect before use —do not use if product or packaging is damaged.
• Do not allow the stylet to protrude beyond the distal end of the ET tube.
• Do not allow the stylet to advance into the glottis or past the vocal cords.
• Single use only. Do not reuse, reprocess, or resterilize.
Instructions for Use
1.
Read all instructions before using the stylet.
2.
Remove the stylet from its protective package.
3.
Load a standard ET tube onto the stylet.
4.
Slide the stylet into the tube until the distal end of the stylet is just inside the distal end of the tube.
5.
Slide the stopper into position in the tube, so that the distal end of the stylet cannot pass the distal end of the tube.
Distal end of stylet
within tube
6.
Position the distal end of the ET tube visually above the vocal cords. Do not advance the stylet past the vocal cords.
7.
When the ET tube is engaged in the glottic opening, use your thumb to disengage the stopper and retract the stylet.
8.
Position the ET tube per standard practice.
9.
Remove the stylet from the ET tube.
10. Examine the stylet for damage, and then dispose of it.
Storage Ranges
Temperature
-20–45 °C (-4–113 °F)

Symbol Definitions

Warning
Refer to Operations and
Maintenance Manual
Manufacturer
Date of manufacture
Use by (date)
Sterilized through radiation
Slide into
position
Humidity
0–95%
Catalogue (part) number
Batch code
Temperature limit
Humidity limit
Atmospheric pressure limit
Atmospheric Pressure
video
®
44–106 kPa
Reuse not allowed
Statement of prescription
Do not use if packaging is
damaged
In accordance with CE
Medical Device Directive
Representative for
European Community