SEITE 2 - 148 x 210 mm - 086/0/-/51+ F -
- to plug or clog the air outlet opening inside the
polymerization chamber (may produce a smear
layer),
- to bypass the lid contact switch (may lead to
insufficient illumination and vacuum),
- to place objects inside the polymerization chamber
which are not specifically intended for this purpose
(may damage the unit),
- to open the lid by force, e.g. with a screw-driver
(may damage the unit).
2. Product Description
Visio Beta vario, manufactured by 3M ESPE, is an
illumination unit for the polymerization of composite veneer
Polymerization
chamber
3. Technical data
Supply voltage:
Power consumption:
Dimensions:
Weight:
Ambient temperature range: 18°C/64°F...40°C/104°F
Relative humidity of air:
4. Start-up of the unit
ATTENTION
Do not start-up the unit plus vacuum pump before the
pump has equilibrated at least to room temperature.
2
SCHWARZ
- 03-0132 (sr)
120 V, 230 V ±10%
50/60 Hz
pre-set voltage and
AC-frequency are stated on
the rating plate
approx. 250 W
(incl. vacuum pump)
280 x 360 x 135 mm
(D x W x H)
approx. 6.4 kg
max. 80% at 31°C/88°F
max. 50% at 40°C/104°F
materials and light-curable dental materials, such as light-
curable tray materials, modeling resins for custom-made
attachment parts or block-out materials.There are four
programs with different vacuum periods available which
can be selected to suit the specific polymerization
requirements of the material in use. In the final step of the
procedure, in which the material is exposed to a vacuum,
the oxygen inhibition layer commonly observed with
composite materials is eliminated.
For details on all products mentioned below please refer to
the corresponding Instructions for Use.
Operating Instructions should not be discarded for the
duration of product use.
+VAC
P1 15 min 15 min
P2 7min 10 sec 10 sec
P3 2 min 2 min
P4 4 min
4 min 4 min
P1
P2
P3
P4
VAC
Power switch
Operating the unit with a cold pump may lead to over-
loading of the pump and damage the pump due to poor
lubrication!
For details on the installation of the Visio Beta Vacuum
pump, manufactured for 3M ESPE, please refer to separate
operating instructions!
Push the suction hose of the Visio Beta Vacuum pump
(labeled "Gerät-unit") onto the hose coupling piece of the
Visio Beta vario unit.
Plug the power cord of the vacuum pump into the upper
socket on the back of the Visio Beta vario unit. Connect one
plug of the power cord to the lower socket on the back of
the Visio Beta vario unit, and the other plug of the cord to a
properly grounded power outlet.
5. Operation
Turn on the power switch. For a total of 3 sec the display
shows the number of times the lamps have been activated
(two digits show the lamps acts in units of thousands/
hundreds) and both lamp error indication lights are on.
Button area:
Program buttons P1-P4
Error indicator lights
Upper lamps
Vacuum
Lower lamps
Display
Start/Stopp button
Start
Stopp
Lid button