Electrical Data - PEWA Benning MM7 Instrucciones De Servicio

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 43
IP 30 means: Protection against access to dangerous parts and protection
against solid impurities of a diameter > 2.5 mm, (3 - first index). No
protection against water, (0 - second index).
-
Operating temperature and relative humidity:
At operating temperature of 0 °C to 30 °C: relative humidity under 80 %.
At operating temperature of 30 °C to 40 °C: relative humidity under 75 %.
-
At operating temperature of 40 °C to 50 °C: relative humidity under 45 %.
-
Storage temperature: The BENNING MM 7 can be stored at temperatures
from - 20 °C to + 60 °C (humidity 0 up to 80 %). The batteries must be
removed from the unit.
7.

Electrical data

Note: The measurement accuracy is stated as the sum of
-
a relative proportion of the measurement value and
-
a number of digits (i.e. numerical steps of the last place).
This measurement accuracy applies for a temperature of 18 °C to 28 °C and a
maximum relative humidity of max. 80 %.
7.1 DC voltage ranges
The input resistance is 10 MΩ (in 400 mV-range 1 GΩ).
Measuring range Resolution
400 mV
4 V
40 V
400 V
1000 V
7.2 AC voltage ranges
The input resistance is 10 MΩ parallel 100 pF. The measurement value is
obtained as a true RMS value and displayed as such. With non-sinusoidal
curves, the value displayed is less accurate. This results in an additional error
for the following crest factors:
Crest factor of 1.4 to 3.0, additional error + 1.5 %.
Crest factor of 3.0 to 4.0, additional error + 3.0 %.
Measuring
Resolution
range
400 mV
100 µV
4 V
1 mV
40 V
10 mV
400 V
100 mV
750 V
1 V
*1
± (1.5 % + 5 Digit) in frequency range 500 Hz - 1 kHz
*2
± (1.5 % + 5 Digit) for measuring ranges > 50 % of the measuring range value
7.3 DC ranges
Overload protection:
-
1 A (500 V) fuse, rapid on mA input
-
10 A (500 V) fuse, rapid on 10 A input
Measuring range Resolution
40 mA
400 mA
10 A
7.4 AC ranges
The measurement reading is obtained as a true RMS reading and displayed
as such. With non-sinusoidal curves, the value displayed is less accurate. This
results in an additional error for the following crest factors:
Crest factor of 1.4 to 3.0, additional error + 1.5 %.
Crest factor of 3.0 to 4.0, additional error + 3.0 %.
Overload protection:
-
1 A (500 V) fuse, rapid on mA input
-
10 A (500 V) fuse, rapid on 10 A input
10/ 2004
100 µV
± (0.25 % of reading + 5 digits)
1 mV
± (0.4 % of reading + 1 digit)
10 mV
± (0.25 % of reading + 1 digit)
100 mV
± (0.25 % of reading + 1 digit)
1 V
± (0.25 % of reading + 1 digit)
Measurement accuracy in
frequency range 40 Hz - 1000 Hz
± (2.0 % of reading + 8 digits) in frequency range
50 Hz - 60 Hz for 400 mV-range
± (1.3 % of reading + 5 digits)*1*2
± (1.3 % of reading + 5 digits)*2
± (1.3 % of reading + 5 digits)*2
± (1.3 % of reading + 5 digits)*2
10 µA
± (0.6 % of reading + 2 digits)
100 µA
± (0.7 % of reading + 2 digis)
10 mA
± (1.0 % of reading + 3 digits)
BENNING MM 7
Measurement accuracy
Measurement accuracy
Overload protection
1000 V
DC
1000 V
DC
1000 V
DC
1000 V
DC
1000 V
DC
Overload
protection
750 V
eff
750 V
eff
750 V
eff
750 V
eff
750 V
eff
Voltage drop
200 mV max.
2 V max.
2 V max.
14

Publicidad

Tabla de contenido
loading

Tabla de contenido