Ohio Medical products have unit serial numbers with coded logic which indicates
a product group code, the year of manufacture and a sequential unit number for
identification.
AAA A 12345
This alpha character indicates the year of product manufacture
and when the serial number was assigned; "Y" = 1995,
"Z" = 1996, "A" = 1997, etc. "I" and "O" are not used.
Warnings - Possible injury to patient or operator
WARNINGS
Never use any petroleum based lubricants in an Oxygen environment,
as these materials are highly combustible in the presence of Oxygen.
The only Oxygen service lubricants recommended for this equipment
are Sentinel
300).
Do not use a flowmeter with any cracked or damaged plastic
components, gas leaks, loose fittings, or knobs or any missing
components. Supply gas pressure can cause parts to be expelled and
injury may occur.
The flowmeter is only intended for the gas specified on the flowmeter label. Do
not use the flowmeter with any other gases. Inaccurate flow indications
and patient injury may occur.
Do not allow oxygen or oxygen equipment to be exposed to fire,
sparks, or electrical equipment which may provide a source of
ignition. DO NOT SMOKE IN THE AREA WHERE OXYGEN IS IN USE.
An Ohio Medical power outlet extension, twin "Y" adapter or extension
adapter must be directly connected to the flowmeter manifold prior to
connection to the supply gas.
.⚠ The Pre-Use Checkout Procedure must be performed before using
this equipment on each patient. If the flowmeter fails any part of the
Pre-Use Checkout Procedure, it must be removed from service and
repaired by qualified service personnel.
After patient use, oxygen therapy equipment may be contaminated. Handle
in accordance with your hospital's infection control policy.
Clean and disinfect all respiratory equipment before shipment or service to
ensure transportation personnel and/or service personnel are not exposed
to any hazardous contamination.
On models with a power outlet, the supply pressure at the flowmeter will
decrease during periods of high flow from the power outlet. This will cause
a decrease in flowmeter accuracy. The actual flow from the flowmeter outlet
will be lower than indicated. For example, with 150 L/min flowing from the
power outlet, the supply pressure drops from 50 psi to 46 psi and the actual
flow from the flowmeter is then lower than indicated by up to 20%.
6700-0363-000
(Rev. 10) 08/2020

Precautions

OPG (6700-0067-200) or Vac Kote
®
37951M (0220-0091-
®