Operation; Instructions For Use; Definitions/Abbreviations; Flowmeter Identification - Ohio Medical 6700 Serie Instrucciones De Uso

Flujómetro de presión compensada
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 67

Definitions/Abbreviations

= Maximum
MAX
= Do not use petroleum based lubricants on this equipment
= Read scale at the ball center line for L/min
FLUSH = > 15 - 90 L/min
L/min or L/min
°C
Degrees Celsius
°F
Degrees Fahrenheit
NPTF
National Pipe
Thread Female (USA)

Flowmeter Identification

1
Flowtube Scale
Ball
1/8" NPTF Inlet Extension
Probe/Adapter Port
Flow Control Knob
Fitting Port
Tubing Nipple (Nut & Gland)
"Y" Adapter

Environmental Specifications

Storage temperature range:
The following tables show the effect of temperature extremes on operating accuracy at
various flow rate settings. Data shows the additional percent offset at given readings:
Pressure Compensated Flowmeter (0-15 L/min range)
Flow rate
1 L/min
3 L/min
5 L/min
10 L/min
15 L/min
6700-0363-000
(Rev. 10) 08/2020

Operation

liters per minute
Temperature
0°C
15°C
(32ºF)
(59ºF)
6 %
4 %
6 %
3 %
4 %
1 %
5 %
0 %
5 %
1 %
Warning, Caution and Attention,
see "Instructions for Use"
Serial Number
SN
Manufacturer
-20°C (-4°F) to +60°C (+140°F)
21°C
40°C
(69.8ºF)
(104ºF)
0 %
13 %
0 %
4 %
0 %
3 %
0 %
4 %
0 %
4 %
FLUSH
FLUSH
FLUSH
0
0
0
2
2
2
15
15
15
15
10
10
10
10
5
5
5
5
4
4
4
4
3
3
3
3
2
2
2
2
1
1
1
1
5