Descargar Imprimir esta página

Satelec Acteon Piezotome 2 Manual De Uso página 5

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 12
b) Pre-disinfection / cleaning of the tip
Pre-disinfection / chemical cleaning
To be carried out immediately after treatment,
wearing gloves.
Immerse the tip in an ultrasonic bath (enzymatic,
quaternary ammonium base detergent / disinfectant
solution), carefully respecting the concentration and
contact time prescribed by the manufacturer of the
solution.
Use a solution that bears the EC marking or that is in
compliance with any standard required under the
applicable national regulations.
Rinse under running water for at least 30 seconds.
Mechanical / chemical cleaning
To be carried out immediately after pre-
disinfection/chemical cleaning, wearing gloves.
Scrub the tip in a new bath containing the same
enzymatic, quaternary ammonium base detergent /
disinfectant solution, using a metallic brush, for at
least 30 seconds, and in any case until all remaining
traces of contamination have been removed.
Rinse again under running water for at least 30
seconds.
c) Drying
Dry with a clean, non-woven single-use cloth, so as to
remove any traces of liquid.
d) Sterilization conditions
Place in single-use sterilization pouches or bags in
compliance with the specifications defined in
standard EN ISO 11607-1, or in any equivalent
standard required by the national regulations.
It is essential to sterilize the handpiece, with the tip
removed.
e) Sterilisation
The P
LED handpiece body and its three
IEZOTOME
elements must be sterilized individually in an
autoclave (Fig. 2-B) under the following conditions:
- Autoclave: Type B compliant with standard EN
13060.
- Sterilization temperature: 134°C.
- Sterilization stabilization time: 18 minutes.
- Pressure: 2 Bars min.
Remove the P
IEZOTOME
from the autoclave as soon as sterilization is
complete.
Caution:
• Always sterilize the P
assembly without the insert.
• After multiple use, and during successive
sterilizations, make sure before use that the
P
LED handpiece-cord assembly has not been
IEZOTOME
damaged on the outside.
f) Assembly
- Connect the LED ring to the P
making sure it is positioned correctly (align the "+"
marks on the LED ring and on the handpiece body
near the LED ring electrical contacts) (Fig. 3)
- Insert the optical guide into the P
handpiece nose.
- Screw on the front part (nose) of the handpiece.
g) Disposal and recycling
Disposal of the device and its accessories should be
via a collection centre for electronic apparatus or by
returning them to Satelec After Sales Service. In no
circumstances should the device and its accessories
be disposed of by the user. The use of the WEEE
symbol
indicates that this product must not be
treated as household waste, but must be treated
separately, in conformity to the Directive
2002/96/CE. By ensuring this product is disposed of
correctly, you will help to protect the environment.
For more detailed information about the recycling of
this product, please contact your local authority, your
house waste disposal service provider or where you
purchased the product. At the end of their life, tips
should be discarded in safety containers designed for
this purpose
5
LED handpiece-cord assembly
LED handpiece-cord
IEZOTOME
LED handpiece,
IEZOTOME
IEZOTOME
LED

Publicidad

loading