Gossen MetraWatt MAVOWATT Serie Guía De Uso página 47

Analizadores de redes
Tabla de contenido

Publicidad

Description
Amps RMS
Derived from
200mS (10/12
cycles 50/60 Hz)
Aggregated to
selected interval
Amp ½ cycle slide
Cyclic RMS of full
cycle restarted every
½ cycle
Amps DC
Derived from
200mS (10/12
cycles 50/60 Hz)
Aggregated to
selected interval.
RMS Deviation
Subtraction of 1
cycle RMS from
adjacent cycles.
Used for cyclic
waveshape
transient trigger
system.
Maximum magnitude
value of Crest.
Largest Absolute
magnitude of
samples in a ½
cycle. Used for cyclic
waveshape transient
trigger system.
DC of individual
Cycle
GMC-I Messtechnik GmbH
Abbreviation
Wiring Configuration
Irms-a
Any
Irms-b
Irms-c
Irms-d
Icyc-a
Any
Icyc-b
Icyc-c
Icyc-d
Idc-a
Any
Idc-b
Idc-c
Idc-d
Icycw-a
Single phase
Icycw-b
Split phase
Icycw-c
Wye
Icycw-d
Ipk-a
Single phase
Ipk-b
Split phase
Ipk-c
Wye
Ipk-d
Icycdc-a
Any
Icycdc-b
Icycdc-c
Icycdc-d
MAVOWATT 230, 240, 270 & 270-400
Formula
where n=512 samples
where n=512 samples
where n=512 samples
Irms(cycle 1) -
Irms(cycle 2)
Largest Absolute
magnitude of 256
samples (1/2 cycle)
where n=512 samples
Units
Precision
Amps
+/- 0.1 % of
Reading
+/- 0.05 %
of FS
9 KHz BW
Amps
+/- 0.1 % of
Reading
+/- 0.1 % of
FS
9 KHz BW
Amps
+/- 0.2 % of
Reading
+/- 0.1 % of
FS
Amps
+/- 0.2 % of
Reading
+/- 0.1 % of
FS
9 KHz BW
Amps
+/- 0.2 % of
Reading
+/- 0.1 % of
FS
Amps
+/- 0.2 % of
Reading
47