Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 72
Push
to take a sample measurement. During measurements, the
parameter flashes and the timer shows the stabilization time. To change
the measurement parameter (if applicable), push and hold
To measure the sample continuously, push
flashes to indicate the continuous measurement mode.
Change the temperature units
The temperature units can be changed when the measurement screen is
shown.
1. Make sure that a stable measurement reading is shown.
2. Push
and
at the same time. The temperature units change to °C
or °F.
Disable the auto-shutoff option
The meter is set to automatically power off after 5 minutes of inactivity to
maximize the battery life. This option can be temporarily disabled.
1. Make sure that the meter is powered off.
2. Push and hold
until the display shows OFF and no. The power
remains on indefinitely.
3. To power off the meter, push and hold
OFF.
Note: The auto-shutoff option is enabled again when the meter is powered on
normally.
Set the backlight time
The display is illuminated when
length of time that the backlight stays on.
Note: The battery life is decreased when the backlight time is increased.
1. Make sure that the instrument is powered on.
2. Push
and immediately push
flashes.
12 English
.
2 times. The parameter
until the display shows
is pushed. The user can set the
2 times. The backlight time
3. Push
or
to change the backlight time (range: 3 seconds to
2 minutes).

Data logger

Set the date and time
Each data point is saved with the time and date of the measurement.
1. From the standby screen, push
2. When the display shows rt, push
date (DD-MM-YY).
3. When the values flash, use the arrow keys to set the current year,
month and day. After the date is set, the display shows tin and the
time (HH:MM:SS).
4. When the values flash, use the arrow keys to set the current hour,
minute and second.
Save data manually
To save data manually, push
any time during a continuous measurement. The display shows DL and
OK each time a data point is saved.
Save data automatically
When the instrument is set to measure continuously, data can be saved
at regular time intervals.
1. From the standby screen, push
2. When the display shows Fr, push
OFF.
3. Push
. The time interval in HH:MM:SS is shown.
4. Use the arrow keys to set the interval when data will be automatically
saved (range: 5 s to 1 min 30 s).
5. Start a continuous measurement. Data will be saved at the specified
interval. The display shows DL and OK each time a data point is
saved.
.
. The display shows dAt and the
when a measurement is complete or at
.
. The display shows Fr and

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Dl

Tabla de contenido