CLEANING, SANITIZATION AND DISINFECTION
Turn off the unit before any cleaning and disconnect the power cord from the mains.
We recommend a personal use of both the nebulizer and accessories to avoid any risk of
infection or contagion.
CLEANING THE DEVICE AND THE TUBE'S OUTER SURFACE
Only use a cloth dampened with antibacterial detergent (non-abrasive and free of solvents
of any kind).
SANITIZATION AND DISINFECTION OF THE ACCESSORIES
Open the nebulizer by rotating the upper part (C5.1) counter-clockwise, detach the nozzle
by pressing as indicated by the 2 arrows in the "Assembly diagram" section C5.
SANITIZATION
Before and after each use, sanitize the nebulizer and accessories according to one of the
methods descrive below.
(Method A): The accessories that can be sanitized with the method A are (C1-C2-
C3-C4-C5.1-C5.2-C5.3-C5.4-C6). Sanitize the accessories in warm (approximately 40°C)
drinking water with mild dishwashing liquid (non abrasive) or in the dishwasher with hot
cycle.
(Method B): The accessories that can be sanitized with the method B are (C1-C2-C3-C4-
C5.1-C5.2-C5.3-C5.4-C6). Sanitize the accessories by immersion in a solution consisting of
50% water and 50% white vinegar. Finally rinse thoroughly with warm (approximately 40°
C) drinking water.
After sanitization shake the accessories vigorously and place them on a paper towel, or
alternatively, dry them with a jet of hot air (e.g. using a hair dryer).
DISINFECTION
The accessories that can be disinfected are (C1-C2-C3-C4-C5.1-C5.2-C5.3-C5.4-C6)
The disinfection procedure described in this paragraph is to be performed before using
the accessories and is effective only if all the prescriptions herein are met and all the
components involved are previously sanitized. The disinfectant to be used must be of
electrolytic oxidizing chlorine type, (active ingredient: sodium hypochlorite), specific for
disinfection and is available in all pharmacies.
Procedure:
- Fill a container, suitable to contain all the individual components to be disinfected, with
a solution of drinking water and disinfectant, while respecting the proportions indicated
on the packaging of the disinfectant itself.
- Completely immerse each individual component in the solution, taking care to avoid
the formation of air bubbles in contact with the components. Leave the components
immersed for the period of time indicated on the packaging of the disinfectant, and
on
associated with the concentration chosen to prepare the solution.
or
- Remove the components now disinfected and rinse thoroughly with lukewarm drinking
he
water.
- After disinfection shake the accessories vigorously and place them on a paper towel, or
ble
alternatively, dry them with a jet of hot air (e.g. using a hair dryer).
ur
- Dispose of the solution as recommended by the disinfectant manufacturer.