MAINTENANCE MODE
Software version and span switch state check.
OPERATION
+
+
+
SPAN ADJUSTMENT
(It can work only when SPAN SWITCH is on Enable)
Full capacity weight for SPAN Adjustment
OPERATION
+
+
+
Ensure no weight on the pan
Press
Place the calibration weight on the pan
(e.g. 3kg)
Put full capacity weight on the pan (e.g
6kg) and then press
Calibration completed
1 - ZERO
2 - NET
DISPLAY 1 2
0.0 0 0
8 8 8 8 8 8
V r x. x x
S - O N
0.0 0 0
1 - ZERO
2 - NET
DISPLAY 1 2
0.0 0 0
8 8 8 8 8 8
C A L 0 0
- - - - - -
C A L S P
6.0 0 0
- - - - - -
key.
6.0 0 0
REMARKS
Weighing mode .
Keep pressed the
key and press
twice the key
.
Release the
key
Display version for 2 seconds.
Display current Span Switch state
(S-ON ó S-OFF)
Exit after 3s
REMARKS
Weighing mode.
Keep pressed the
and press the keys
+
and
Release the
key
Calibrating zero point.
Displays "CALSP" and full capacity (e.g. 6 kg)
alternately.
Span calibration
The balance goes back to weighing mode.
8