Chapter 8
TROUBLE SHOOTING
Trouble
1. The SpO
2
and Pulse Rate
can not be
displayed normally.
2. The SpO
2
and Pulse Rate
are not displayed
stably.
3. ECG display
astatically.
4. The device can
not be turned on.
5. Noise.
6. The device can
not be used for full
time after charge
or the battery can
not be full charged
even after 10 hours
charging time.
Note:
 If the trouble is still not got rid of, please contact warrant service agent
as soon as possible.
 For the damage aroused by client dismantling and mending, our company
does not assume any responsibility.
Possible Reason
1.The finger is not
properly positioned.
2.The patient's SpO
value's going beyond
the limits.
1. The finger is not placed
inside deep enough.
2. The finger is shaking
or the patient is moving.
1. Electrodes do not
connect with skin very
well.
2.The device is shaking.
3. The patient is shaking
or moving.
1. The battery is drained
away or almost drained
away.
2. The malfunction
of the device.
Here is disturbance
sources.
The battery is broken.
ENGLISH
Solution
1. Place the finger properly
and try again.
2.Try again. Go to a hospital
2
for a diagnosis if you are sure
the device works all right.
1. Place the finger properly
and try again.
2. Let the patient keep still.
1. Recoat gel and press
electrodes.
2.Keep the device on the
body still.
3. keep the patient in a stable
state.
1. Recharge battery.
2. Please contact the local
service center.
Remove the disturbance sources
and other electrical instruments.
Please contact the local
service center.