Descargar Imprimir esta página

Set Up The Calculation; Modbus Registers - Hach 4-20 mA Manual Del Usuario

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 22
5. Select SET FUNCTION>LOGARITHMIC and then select the
applicable options in the ACTIVATION menu.
Option
Description
SET 50% VALUE
Sets the value corresponding to 50% of the process
variable range.
SET HIGH VALUE Sets the high endpoint (upper value) of the process
variable range.
6. Select SET FUNCTION>BILINEAR and then select the applicable
options in the ACTIVATION menu.
Option
Description
SET LOW VALUE
Sets the low endpoint of the process variable
range.
SET HIGH VALUE
Sets the high endpoint (upper value) of the
process variable range.
SET KNEE POINT
Sets the value at which the process variable
VALUE
range divides into another linear segment.
SET KNEE POINT
Sets the value of the current at the knee point
CURRENT
value.
Set the error hold mode
1. Select SETUP SYSTEM>SETUP OUTPUTS>ERROR HOLD MODE.
2. Select an option.
Option
Description
HOLD OUTPUTS
Holds the outputs at the last known value when
communications are lost.
TRANSFER
Switches to the transfer mode when communications
OUTPUTS
are lost. The outputs transfer to a pre-defined value.

Set up the calculation

Set up variables, parameters, units and formulas for the analyzer.
1. Select CALCULATION.
2. Select an option.
Option
Description
SET VARIABLE X
Selects the sensor referring to the variable X.
SET PARAMETER X Selects the parameter referring to the variable X.
SET VARIABLE Y
Selects the sensor referring to the variable Y.
SET PARAMETER Y Selects the parameter referring to the variable Y.
SET FORMULA
Selects the calculation formula to complete. Options:
None, X-Y, X+Y, X/Y, [X/Y]%, [X+Y]/2, X*Y, [X-Y]%/X
DISPLAY FORMAT
Selects the number of decimal places shown in a
calculation result. Options: Auto, XXXXX, XXXX.X,
XXX.XX, XX.XXX, X.XXXX
SET UNITS
Enters the unit name (maximum of 5 characters).
SET PARAMETER
Enters the measurement name (maximum of
5 characters).

Modbus registers

A list of Modbus registers is available for network communication. Refer
to www.hach.com or www.hach-lange.com for more information.
English 11

Publicidad

loading