PREDICTED VALUE SET UP

To change the Predicted Value Settings, from the default shown on the
box, hold down the On
release when 'Prd' is displayed. Use the Scrolling Up
keys to choose an alternative Predicted Value set.
4
1 ECCS
5
2 NHANES III
3 Asian (Chinese)
When the desired number is selected press the Enter
accept and save that predicted value set as default.
To check the version number, hold down the On
Down
keys together, and only release when 'rE' is displayed. The
version number will then be shown.

TROUBLE SHOOTING INFORMATION

Should you encounter problems operating your Pulmolife unit, please
consult the table below:
Problem
Does not register a
blow
Unit does not turn on
or Solid "bAt"
indicator is displayed,
while unit beeps
three times and
powers itself down
after 5 seconds
Calibration Check
failed or cannot be
completed
English
and Enter
6
VERSION
Possible
Cause
Turbine
incorrectly
fitted
Battery is
exhausted
There are
leaks in the
syringe or
connections
Turbine may
be faulty
Shaft of the
syringe being
pushed down
keys together, and only
/ Down
key to
and Scrolling
Solution
Ensure turbine is
tightly pushed into
the unit.
Replace battery
Check the syringe
and connections for
leaks
Repeat calibration
procedure, if
problem still
persists, replace
turbine
The syringe should
be emptied and
filled with one
smooth stroke,
avoid pushing down
on the shaft or
7