DATA
This feature is to store the data measuring under infrared mode.
1. The maximum number of data to be stored is 99.
2. When the storage is empty, it displays '--- . ---'
3. Under non-measurement mode, press and hold DATA and
data, the Thermometer beeps.
4. Under measurement mode, press YELLOW button to toggle to DATA mode.
5. Press
or
to select the location of storing the data.
6. Press DATA to store the value.
TC-K THERMOCOUPLE MEASUREMENT
This feature is to measure temperature with K type thermocouple.
1. Press YELLOW button to toggle to T-C mode. The LCD display shows OL before the
thermocouple is connected.
2. Connect the K type thermocouple, press the trigger to start measuring. The LCD display shows
the reading.
3. Release the trigger, the reading is kept, LCD shows T-C. The temperature reading is shown on
the bottom of right hand side.
HOLD
The display will remain activated 8 seconds after the trigger is released. HOLD appears in
the upper middle of the display. When the trigger is pulled again, the Thermometer will begin
measuring in the last function selected.
This section describes a variety of measurements often performed by technicians.
Reminder:
- User could select to turn on or off the backlight and laser whenever you are making readings
with the Thermometer. But if you are using USB to power up the Thermometer, the two levels
white colour backlight will be on automatically.
- Relatively high emissivity normally means emissivity setting of about 0.95.
- Relatively low emissivity normally means emissivity setting of about 0.30.
- When user cannot identify the emissivity of the object to be measured, user could cover the
surface to be measured (temperature >150 ) with black electric tape (emissivity of about 0.95).
Allow time for the tape to reach the same temperature as the object to be measured. Measure
and record the temperature of the tape.
Target the Thermometer to the object to be measured, adjust the emissivity setting to make it as
the same temperature as the tape. At this time, the Thermometer emissivity setting is close to the
emissivity of the object to be measured, measurement could be started.