Operation
TO ADJUST CALIBRATION
IMPORTANT: Always obtain three (3)
test results from the grain elevator for
the sample being compared. Average
these three (3) results. Compare this
average with the average of three (3)
tests by the moisture tester.
1. Each grain scale can be individually
adjusted up to 5.0% by increments of
0.1% to more closely match the results
of an elevator tester.
2. First select the grain to be adjusted.
3. Test the grain selected, using the
sample you wish to adjust to. Once
valid moisture appears, press the CAL button (A).
NOTE: Calibration of given grain cannot be performed unless a valid mois-
ture test has been performed.
4. Once CAL button (A) is pressed, tester will now display the obtained mois-
ture reading and the current offset applied to that moisture range.
5. Press the UP arrow (B) to raise the adjustment amount or press the DOWN
arrow (C) to lower. The tester will add or subtract up to 5.0% by increments
of 0.1% to the current moisture range.
6. After the adjustment amount has been selected, press the CAL button (A)
to return to grain tester mode.
NOTE: Adjustment can NOT be made to factory calibration to produce
moisture readings in a tester that shows "BELOW LIMIT" or "ABOVE LIMIT"
readings.
IMPORTANT:
This tester incorporates Multi-Point Calibration for each grain. Therefore,
once a valid test is taken and an adjustment is made, the adjustment will
only affect the moisture range of the sample tested.