Operating Procedure; Switching On; Dc Zero Adjustment; Measurement - Chauvin Arnoux PAC 10 Manual De Instrucciones

Pinza amperimetrica
Ocultar thumbs Ver también para PAC 10:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

3.1/ SWITCHING ON

Set the sliding switch
Correct operation is indicated by a green light
is in good condition.
If this green indicator does not come on when the clamp is switched on,
or goes out during the operation of the clamp, it is necessary to replace the
battery (see MAINTENANCE chapter).
3.2/ ZERO ADJUSTMENT
Ensure that the jaws
enclose any conductor. Connect the clamp to your measurement instrument
set up according to the appropriate voltage range then turn the rotative button
until zero DC is displayed on the measurement instrument

3.3/ MEASUREMENT

After having switched on the clamp, connected it to the measurement
instrument on the appropriate range, and set the zero DC procedure (see
the two paragraphs above), clamp the conductor to be measured
between the jaws of the clamp.
The value is displayed on the measurement instrument.
Depending on the sensitivity selected on the clamp and the range of your
measurement instrument, apply the conversion ratio to get the value of the
current.
On DC current measurement, ensure that the arrow located on the
external edge of the jaws
current flowing in the conductor. (source Þ receiver)
Ranges
400 A
4.1/ RÉFERENCE CONDITIONS
- Temperature: 18...28°C
- Humidity rate: 20...75% RH
- Battery voltage: 9 V ± 0.1 V
- Position of conductor: centred on the markings of the clamp
- Magnetic field: Earth's DC field
- Absence of external AC magnetic field
- Absence of electric field
- Measurement for a DC current or an AC sinusoidal current £ 65 Hz
- Impedance of the measurement instrument : ³ 1 MW and £ 100 pF

4.2/ OPERATING CONDITIONS

The instrument must be used in the following conditions to satisfy the
safety of the user and the metrological performance:
- Use indoors
- Working altitude:£ 2000 m
- Transportation altitude:£ 12 000 m
- Environmental conditions: see graph below

3/ OPERATING PROCEDURE

to the position 400 A range (sensitivity 1 mV/A).
of the clamp are correctly closed and that they do not

4/ SPECIFICATIONS

input/output
ratio
1 mV/A
corresponds to the direction of the
measurement extent
AC rms
0.5...400 A
10
indicating that the battery
peak max
0.5...600 A
0.5...600 A
DC

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

1200.70

Tabla de contenido