iCH Prime/English
Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
Flicker emissions
IEC61000-3-3
Warning:
1. The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the device should be observed to verify normal operation in the configuration in which it will be used.
2.Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
3. Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the CPAP,including
cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Basic EMC standard
Electrostatic Discharge
(ESD) IEC61000-4-2
Electrical fast transient/
burst
IEC61000-4-4
Appendix A: EMC Information
Compliance
Electromagnetic Environment-Guidance
The device uses RF energy only for its internal function.
Group1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment
Class B
The device is suitable for use in all establishments, including
Class A
domestic establishments and those directly connected to the
public low-voltage power supply network
Complies
Immunity Test Levels
Professional
healthcare
facility
environment
±8kV contact
±15kV air
±2kV for power supply line
±1kV for input/output line
Compliance
HOME
Levels
HEALTHCARE
ENVIRONMEN
T
±8kV contact
±15kV air
±2kV for power
supply line
±1kV for
input/output line
Electromagnetic
Environment-Guida
nce
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be that of atypical
commercial or hospital
environment
22