sor. Correct filter maintenance prolongs
the aerosol's life. Filter replacement ti-
mes depend on the external conditions of
use of the device. We recommend repla-
cing filters after around 20 hours of use
or if they are visibly dirty.
The filter is on the right side of the aero-
sol unit (c).
To replace the filter :
• Open the filter cover
• Remove the filter from its housing
• Carefully fit the new filter
• Re-fit the filter cover
TECHNICAL DATA
Nominal voltage 230 V
Frequency 50Hz
Power: 180VA
Fuse: T 5A 250V ~
Maximum flow rate : 17 l/min 2
Operating flow : 7 l/min 1
Maximum pressure : 2,6 0,2 atm
Working pressure : 0,8 0,2 atm
Protection against liquid seepage : IPX0
Weight:1,7Kg
Conditions for using the device:
20 ON – 40 OFF
Tmax=35°C maximum ambient operating
temperature for the device to prevent it
from overheating
MMAD: 2,65 μm 0,11
Breathable part :
79% 1,72
Nebulizer cup capacity : 5 ml
Noise level at 50 cm* : 68,0 dB
*Data on noise levels were measured on
a new device. Values may vary with use.
The performance information supplied
by the manufacturer in accordance with
standard EN 13544-1 may not apply to
drugs in suspension or highly viscous
drugs.
EN 60601-1 Medical electrical equipment
– Part 1 : general requirements for sa-
fety
EN 60601-1-2 Medical electrical equip-
ment – Part 1 : general requirements for
safety – Collateral standard : electroma-
gnetic compatibility – Requirements and
tests
LEGEND OF
SYMBOLS:
Caution! Read instructions
Type B device
Double insulation
Complies with directive
MDD 93/42/EEC
Alternate current
Conditions for storage :
Relative humidity: 45% / 85%
Temperatures: -20°C / +70°C
"The medical device requires special pre-
cautions regarding electromagnetic com-
patibility and shall be installed and put
into service according to the information
provided in the user manual"