10.0
Immunity Test
Voltage Dips / Dropout
IEC 61000-4-11
NOTE
U
is the mains alternating current prior to application of the test levels.
T
10.3 Guidelines and Manufacturer´s Declaration - Immunity
The ATMOS Patient chair M2 is intended for use in the electromagnetic environment specifi ed below. The customer
or user of the ATMOS C161 should ensure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
30
IEC 60601-
Test Level
< 5 % U
T
(> 95 % Dip of the U
)
T
for 0.5 Cycle
40 % U
T
(60% Dip of the U
)
T
for 5 Cycles
70% U
T
(30 % Dip of the U
)
T
for 25 Cycles
< 5 % U
T
(>95 % Dip of the U
)
T
for 5 s
IEC 60601-
Compliance Level
Test Level
3 V
3 V
eff
150 kHz to up 80 MHz
3 V/m
80 MHz to up 2,5 GHz
10 V/m
Compliance Level
< 5 % U
T
(> 95 % Dip of the U
)
T
for 0.5 Cycle
40 % U
T
(60% Dip of the U
)
T
for 5 Cycles
70% U
T
(30 % Dip of the U
)
T
for 25 Cycles
< 5 % U
T
(>95 % Dip of the U
)
T
for 5 s
Electromagnetic Environment
- Guidance
Portable and mobile communications
equipment should be separated from the
Patient chair E1 incl. the cables by no less
than the distances calculated/listed below.
Recommended distances:
d = 3,5/3 √ P
d = 3,5/10 √ P
d = 7/10 f √ P
where „P" is the max. power in watts (W)
and D is the recommended separation
distance in meters (m).
Field strengths from fi xed transmitters, as
determined by an electromagnetic site (a)
survey, should be less than the compli-
ance level (b).
Interference may occur in the vicinity of
equipment containing following symbol
Electromagnetic
Environment - Guidance
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
ATMOS Patient chair M2 demands
continued function even in case of
interruptions of the energy supply,
it is recommended to supply the
ATMOS Patient chair M2 from an
uninterruptible current supply or a
battery.