Storage
Keep the device and accessories dry. Avoid exposure to
direct sunlight.
TRoUBLeShooTINg
If your Neb-u-Tyke Speedster Nebulizer Compressor fails
to function, consult the following Troubleshooting guide. If
the problem persists, consult your Graham-Field dealer.
Problem
Device doesn't
operate
Nebulization is
weak
John Bunn Neb-u-Tyke Speedster Nebulizer Compressor
Cause and Solution
Check if plug is properly fit into a grounded electrical outlet.
When device has been run continuously for more than 30 minutes
just before using, auto-shut-down may have been activated by the
built-in thermal protector; ensure that device is OFF for at least 30
minutes before next usage.
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to Neb-u-Tyke
compressor or nebulizer cup; replace as needed.
Check that the nebulizer cup is properly assembled and not
damaged. If there is any damage, replace as needed.
Check if filter is dirty; replace filter as needed.
11