Aesculap Suhl GmbH
Battery-powered clipping machine FAV5 CL
5.
Processing
Damage to, or destruction of the
product
cleaning/disinfecting or steriliza-
tion!
CAUTION
Only clean and disinfect the
product manually.
Do not sterilize the product
under any circumstances.
5.1 Cleaning/disinfection
Note
Always follow disposal and hygiene instructions and
regulations!
Manual cleaning/disinfection
Risk of electric shock and fire haz-
ard!
Unplug the device before clean-
DANGER
ing.
Ensure that no fluids will pene-
trate the product.
Leave the product exposed to
air for at least 1 minute after
cleaning/disinfecting.
10
caused
by
mechanical
Damage to the product due to
incorrect cleaning or inappropriate
cleaning/disinfecting agents!
Use any commercial detergent
CAUTION
for surface cleaning.
Never immerse the product in
water or detergent.
Do not introduce any fluid into
the charging bays.
Clean the contacts in the charg-
ing bays, applying extreme care.
Wipe the shell of the product with a lint-free cloth
moistened with a commercial disinfectant for wipe
disinfection.
If necessary, after the appropriate exposure time,
wipe off any residues of the cleaning/disinfecting
agent with a lint-free cloth moistened with clean
water.
Dry the product with a clean, lint-free cloth.
Clean the contacts in the charging bay with isopro-
panol or ethyl alcohol and a cotton swab. Do not
use corrosive chemicals.
If necessary, repeat the cleaning/disinfecting pro-
cess.
5.2 Control, care and inspection
Allow the product to cool down to room tempera-
ture.
Inspect the product after each cleaning and disin-
fecting cycle to be sure it is: clean, functional, and
undamaged.
Lubricate blade if required, see Chapter Lubricating
the blade head.
Check the product for any damage, abnormal run-
ning noise, overheating or excessive vibration.
Inspect the blade for broken, damaged or blunt
cutting edges.
Set aside the product if it is damaged.