VO?
Reads and returns the actual output voltage in Volts in <nr2> numeric format. The
syntax of the response is
Example:
IO?
Reads and returns the actual output current in Amps in the <nr2> numeric format.
The syntax of the response is
Example:
OUT?
Returns status of the output as
M?
Returns output mode status as
ERR?
Returns the value in the error register in the form ERR <nr1><rmt>
Miscellaneous Commands
*RST
Resets the instrument to the default power-up settings (1.00V, 1.00A, output off).
*IDN?
Returns the instrument identification. The exact response is determined by the
instrument configuration and is of the form <NAME>,<model>, 0, <version><rmt>
where <NAME> is the manufacturer's name, <model> defines the type of
instrument and <version> is the revision level of the software installed.
V<nr2><rmt>
If the actual output voltage is 12.55 Volts the
response to the command VO? will be V12.55<rmt>.
I<nr2><rmt>
If the output current is 0.93 Amps the response to
the command IO? will be A0.93<rmt>.
OUT ON<rmt>
or
OUT OFF<rmt>
M CV<rmt>
or
M CC<rmt>
0 = no errors.
1 = command not recognised.
2 = command value outside instrument limits.
9