Beurer medical FT 55 Instrucciones Para El Uso página 84

3 en 1 termómetro para el oído
Ocultar thumbs Ver también para FT 55:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 31
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The FT55 series is intended for use in the electromagnetic environment specified below. The customer or the user of the FT55 series should
assure that it is used in such an environment.
IMMUNITY test IEC 60601 test
level
Electrostatic
6 kV contact
discharge (ESD)
8 kV air
IEC 61000-4-2
2 kV for power
Electrical fast
supply lines
transient/burst
1 kV for input/
IEC 61000-4-4
output lines
1 kV line(s) to
Surge
line(s)
IEC 61000-4-5
2 kV line(s) to
earth
<5 % UT
(>95 % dip
in UT)
for 0,5 cycle
interruptions
40 % UT
and voltage
(60 % dip in UT)
variations on
for 5 cycles
power supply
input lines
70 % UT
(30 % dip in UT)
IEC 61000-4-11
for 25 cycles
<5 % UT
(>95 % dip
in UT)
for 5 sec
Compliance
Electromagnetic environment – guidance
level
6 kV contact
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material,
8 kV air
the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Not applicable
Mains power quality should be that of a typical commercial or hospital environment.
Not applicable
Portable and mobile RF communications equipment should be used no closer to any part of
the FT55 series, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2
d = 1.2
80 MHz to 800 MHz
Not applicable
d = 2.3
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey, a should be less than the compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked with the following symbol:
84

Publicidad

Tabla de contenido
loading

Tabla de contenido