4.6 Resistance
Resolution
400
100m
1
4000
NB :
Max open circuit voltage : 4,4V
Max short circuit current : 1,2mA
4.7 Continuity sound test
Continuous sound signal for a resistance less than
40 ± 10
Reponse time : 10ms.
Max open circuit voltage : 4,4V.
Max short circuit current : 1,2mA
Permitted accidental overload : 600V rms
4.8 Diode test
Measurement of the voltage in volts (V) forward direction,
OL or "voltage > 3,5V" in reverse direction.
Max open circuit voltage : 4,4V.
Max short circuit current : 1,2mA
Accuracy : 3% R + 10ct, Resolution : 1mV
Permitted accidental overload : 600V rms
4.9 Frequency (for the F13N only)
Hz
Resolution
1Hz
4kHz
Trigger threshold : 1V or 1A.
NB : frequencies higher than 4kHz can be indicated, but only for
information purposes.
4.10 Max value (MAX)
Max. value of the positive signal measured.
This function is active for AC currents and DC and AC
voltages.
Typical acquisition time : < 500ms.
Additional error :
For the F11N :
± 2,5% R ± 3ct (high calibre), ± 2,5% R ± 30ct (low calibre).
For the F13N :
Accuracy of the calibre + additional error : ± 2,5% R ± 5ct.
4.11 Memorisation (HOLD)
This function records the displayed value and is active for
all the switch positions.
Accidental
Accuracy
brief overload
± 1% R ± 5ct
600V rms
± 1% R ± 3ct
Accidental
Accuracy
brief overload
1000V rms or
± 0,1% R
1000A peak
± 1 ct
20