Tabla De Contenido - HTL LABMATE pro LMP2 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 30
CONTENTS
1 - INTRODUCTlON
1 - INTRODUCTION
The LABMATE PRO is a volumetric instrument de signed
to measure and transfer liquids precisely and safely. It can
measure and transfer volumes from 0.1 μl to 10000 μl
depending on the model.
The pipettes are equipped with a digital counter which
shows the pipetting volume. The volume setting is done
by turning the pipetting pushbutton knob (Fig. 1A2) or
the black adjustment knob (Fig. 1B) in the right direction.
The volume range is shown on the pipetting pushbutton
(Fig. 1A1).
The volume range from 0.1 μl to 10000 μl is covered by
10 pipette models.
Model
Volume range [μl]
LMP2
0.1 - 2
LMP10
0.5 - 10
LMP20
2 - 20
LMP50
5 - 50
LMP100
10 - 100
LMP200
20 - 200
LMP250
50 - 250
LMP1000
100 - 1000
LMP5000
500 - 5000
LMP10000
1000 - 10000
1
Measurement and transfer of micro-volumes,
LMP2, LMP10
DNA sequencing and enzyme-assay applications.
Measurement and transfer of general aqueous
LMP20, LMP50, LMP100,
solution, acids and bases
LMP200, LMP250, LMP1000
LMP5000, LMP10000
Measurement and transfer of large volumes.
The liquid is aspirated into disposable tips attached to
the pipette shaft. Disposable tips ensure maximum safety
and eliminate possibility of cross-contamination of the liquid
samples.
To protect the user from contaminated tips the pipette is
equipped with a built-in tip ejector, (Fig.1D).
The construction of the ejector enables the user to adjust
its length. The adjustable tip ejector accommodates
every variety of tips available on the market. When using
narrow tubes, it may be necessary to remove the tip
ejector.
- in 2-1000 μl pipettes
Tip ejector disassembly (Fig. 5A)
1. Press the tip ejector button.
2. Turn the ejector bush left to the very end.
3. Slide the ejector from the plastic arbor.
Tip ejector assembly (Fig. 5B)
1. Press the tip ejector button.
2. Place the tip ejector on the arbor.
3. Turn the plastic ejector bush right until it's latched.
Tip ejector adjustment (Fig. 5C)
1. Remove the ejector pushbutton.
2. To increase the ejector length turn the plastic bush to
the left.
3. To decrease the ejector length turn the plastic bush to
the right.
- in 5000 and 10000 μl pipettes
Tip ejector disassembly (Fig. 6A)
1. Remove the tip ejector button
2. Use the screwdriver to turn the metal arbor left to
unscrew the ejector from the arbor.
3. Release the ejector.
ENGLISH
2

Publicidad

Tabla de contenido
loading

Tabla de contenido