Compressor Cleaning; Filter Change; Maintenance - Medquip MQ6003 Guía De Instrucciones

Ocultar thumbs Ver también para MQ6003:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3. Keep the outer surface of the tubing dust-free by wiping regularly. Nebulizer tubing does
not have to be washed because only filtered air passes through it.
NOTE
The Drive Disposable Nebulizer will last at least 15 days and possibly
longer, depending on usage. Proper cleaning will also help extend the life
of the nebulizer. Because it is disposable, we recommend that an extra
nebulizer be kept on hand at all times. The Drive Reusable Nebulizer is
dishwasher safe and may be cleaned and reused for up to one year.

Compressor Cleaning

1.
With power switch in the "Off " position, unplug power cord from wall outlet.
2.
Wipe outside of the compressor cabinet with a clean damp cloth every few
days to keep dust-free.
Do not submerge in water; doing so will result in compressor
damage.

Filter Change

1.
Filter should be changed every 6 months
or sooner if filter discolors.
2.
Remove filter by grasping it firmly and pulling
out of the unit. Discard old filter.
3.
Replace with a new Drive filter. Additional
filters should be purchased from your
Drive provider.
4.
Push new filter into position.
Do not reuse or wash filter or substitute any other material
such as cotton for a Drive air-inlet filter. It will result in
compressor damage.

Maintenance

All maintenance must be performed by a qualified Drive provider or authorized service
center.
Electric shock hazard. Do not remove compressor cabinet.
All disassembly and maintenance must be done by a qualified
Drive provider.
Rev.2.08.10.15
12
...........................................................................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL MQ6003, MQ6003R, MQ6004, MQ6005, MQ6005R
...........................................................................................................................................................
WARNING!
Filter
cover
WARNING!
DANGER!

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mq6003rMq6004Mq6005Mq6005r

Tabla de contenido