Exposure Switch

The Exposure Switch allows the operator to control image acquisition from outside of
the X-ray room.
Press and hold the Exposure Switch down until acquisition is completed. Premature
release of the Exposure Switch will abort image acquisition.
Pressing the Exposure Switch activates the LED indicator to turn yellow. This color
indicates that the X-ray is being emitted.
Green 16 User Manual
Exposure Switch
X-ray Exposure LED
• Green: Standby(X-ray Off)
• Yellow: In Operation(X-ray On)
The Exposure Switch is detachable. Ensure the Exposure
Switch cable is not detached from the unit accidentally during the
operation.
Keep vocal/visual contact with the patient during exposure. If any
problem occurs during exposure, release the Exposure Switch
immediately.