SETUP AND DRIVE
OPERATION
1. Unpack the drive: save packaging material until proper prod-
uct operation has been verified.
2. Select and mount Pump Head and load tubing
(see Pump Head manual).
3. For 77200-20 plug in Universal Power supply and connect
to the back of the drive. (See Figure #3.)
4. For 77200-12 supply 11.0 to 15V DC to terminal strip
positions 1 and 2. (See Figure #3.)
5. The drive system can be set to operate in either a clockwise
or a counterclockwise direction. The same control used to
select direction also turns power on or off. To
operate in a clockwise direction place direction control
switch in the "up" position. The POWER indicator should
light and the fluid will flow from right to left. To operate in
a counterclockwise direction, place direction switch in the
"down" position. The POWER indicator should light and
the fluid will flow from left to right. To turn pump off, place
switch in the "center" position.
6. The pump speed is controlled by the variable speed control.
Turning the control clockwise increases the speed. Tubing
life is decreased with increased operating speed. Speed can
be controlled up to a 6:1 range.

MAINTENANCE

Cleaning: Keep the drive enclosure clean with mild
detergents. Never immerse nor use excessive fluid.
Motor/ Brush Replacement: The motor used in this
inexpensive drive is not serviceable or repairable and has
brushes which cannot be replaced. The motor is rated for
a continuous duty life of 3000 operating hours minimum.
Operation at other than continuous duty cycles will result in
longer motor life. If motor failure for whatever reason occurs,
the motor must be replaced.
®

SPECIFICATIONS

Output:
Operating Speed:
Number of Pump Heads:
Torque Load:
Direction of rotation:
Input:
Operating Voltage/Frequency:
Model 77200-20
Model 77200-12
Installation Category:
Model 77200-12 & -20
Remote Start/Stop:
Construction:
Dimensions (L × W × H):
Weight:
Color:
Material:
Enclosure Rating:
Environment:
Operating Temperature:
Storage Temperature:
Humidity (Non-cond):
Altitude:
Pollution Degree:
Compliance (For CE Mark):
EN61326-1/A2: 2001 (EMC Directive)
Converter is UL listed and CSA approved.
Regulatory agency specifications not applicable to the
balance of the unit due to low voltage.
35 to 200 rpm
1
20 oz-in, maximum
Clockwise and Counterclockwise
115V AC nominal, 50/60 Hz
(90–130VAC) @500 mA AC
230V AC nominal, 50/60 Hz
(190–260VAC) @300 mA AC
13.5V DC nominal
(11.0–15.0V DC) @ 2.4 A DC
Category I per IEC664 (Signal Level)
Contact closure connection at
terminal strip contacts 3 and 4
5.750 in × 5.14 in × 4.14 in
(14.6 cm × 13.0 cm × 10.5 cm)
1.88 pounds (0.85 kg)
Black
Painted Steel
IP22 per IEC529
32°F to 104°F (0°C to 40°C)
–49°F to 149°F (–45°C to 65°C)
10% to 90%
6600 ft (2000 m)
Pollution Degree 2 per IEC664
(Indoor usage — lab, office)
®
5