3
Operation
9600 Plus Calibration Tester Setup
Place the 9600 Plus Calibration Tester on a level surface away
from sunlight, drafts, and other sources of heat or cold. The
9600 Plus Calibration Tester takes approximately 20 minutes
to heat to the lowest setting. When testing several
thermometers at all three temperatures, it is recommended
to test all probes at one Calibration Set Point Temperature
before proceeding to the next Calibration Set Point
Temperature. To further expedite testing start at the lowest
Calibration Set Point Temperature. The 9600 Plus Calibration
Tester does not have an internal fan, causing the cool down
time to be greater than the time it takes to heat up.
Changing the Calibration Set Point
Temperature
To scroll from one set point to the next, press and hold the
Temperature Selection button until a beep is heard. The
newly selected set point appears in the upper left corner of
the LCD display. The device's current temperature is
displayed, starts to flash, and continues flashing until the
cavity reaches the equilibrium at the new set point.
13