Select A Different Calibration Mode; Calibration Consideration / Limits - AgraTronix MT-PRO Manual De Instrucciones

Ocultar thumbs Ver también para MT-PRO:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 102

SELECT a dIFFERENT CaLIBRaTION MOdE

1. Calibration mode 'SIMPLE' is the default
mode from the factory.
2. To select a new calibration mode, press either
the up or down arrow on the SELECT button
(A) to index forward or backward through
the grain (function) menu until the phrase
CALIBRATION MODE is displayed.
3. When CALIBRATION MODE is displayed,
press TEST button (B). The current
calibration mode selected will be displayed.
Press either the up or down arrow on the
SELECT button (A) to index forward or
backward through the calibration mode menu until your choice of
calibration mode is displayed. The only 2 mode options are SIMPLE
and ADVANCED.
4. Press TEST button (B) WAIT will be displayed, and then you will be
returned to the Calibration Mode menu.
5. If you do not want to change calibration mode simply press the
CLEAR button and you will be returned to the Calibration Mode menu.
NOTE: CHaNGING CaLIBRaTION MOdES WILL ERaSE aLL
uSER CaLIBRaTION SETTINGS.
CaLIBRaTION
CONSIdERaTION / LIMITS
Typically the moisture tester will be
consistent with most elevator testers
over a broad range of moisture levels.
However, there are some things to
consider when thinking about your
moisture tester calibration.
NOTE: Graph 1 provided is for
illustration purposes only and does
not reflect actual test data.
Operation
05-8
MOISTuRE TESTER
ELEvaTOR TESTER
GRaPH 1:
Moisture Tester - Elevator Tester
Comparison

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido