6.10 Open Dilution Programming
1. Press
repeatedly until the display reads:
2. Press
to enter selection.
The display shows the last used setting for the diluent aspiration volume.
3. Press
to select a new diluent volume (0.10-5.00 mL). (In this case
4.50 mL).
4. Press
to enter selection.
The display shows the last used setting for the sample aspiration volume.
5. Press
to select a new sample volume. (In this case 0.50 mL).
6. Press
to enter selection.
7. Attach the tip to the tip cone.
The pipettor is now ready to begin the dilution. The display indicates that
the pipettor is ready to aspirate the selected diluent volume.
8. Place the tip in the diluent and press the Start-button. The unit will aspirate the diluent volume (4.50 mL) selected.
The following display indicates that the unit is ready to aspirate the
selected sample volume.
9. Place the tip in the sample solution and press the Start-button to aspirate 0.50 mL.