Guidance and manufacturer's declaration - electromagnetic immunity
The 200-series devices are intended for use in the electromagnetic
environment specified below. The customer or the user of a 200-series device
should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
Electrostatic
±6 kV contact
Discharge
±8 kV air
(ESD)
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
±1 kV for input/
IEC 61000-4-4
output lines
Surge
±1 kV
differential
IEC 61000-4-5
mode
±2 kV common
mode
Voltage dips,
<5% U
(>95%
T
short
dip in U
) for
T
interruptions
0,5 cycle
and voltage
variations on
40% U
(60%
T
power supply
dip in U
) for
T
input lines
5 cycles
70% U
(30%
IEC 61000-4-11
T
dip in U
) for
T
25 cycles
<5% U
(>95%
T
dip in U
) for
T
5 sec
Power
3 A/m
frequency (50/
60 Hz)
magnetic field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test level
T
Duo 200
Compliance level Electromagnetic environment
- guidance
±6 kV
Floors should be wood, concrete or
contact /
ceramic tile. If floors are covered
±8 kV air
with synthetic material, the relative
No loss of
humidity must be at least 30%.
performance
±2 kV power / ±1 kV
Mains power quality should be that
I/O
of a typical commercial or hospital
No loss of
environment.
performance
±1 kV diff. / ±2 kV
Mains power quality should be that
comm.
of a typical commercial or hospital
No loss of
environment.
performance
U
- 100% (0,5
Mains power quality should be that
T
period)
of a typical commercial or hospital
No loss of
environment. If the user of a 200-
performance
series device requires continued
operation during power mains
U
- 60% (5 periods)
T
interruptions, it is recommended
No loss of
that the 200-series device be
performance
powered from an uninterruptible
U
- 30%
T
power supply or a battery.
(25 periods)
No loss of
performance
U
- 100%
T
(5 seconds)
Device resets to a
safe state. (60601-1
§ 49.2)
Not applicable
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.