Operation
2. Rinsing
3. Manual
Cleaning
4. Rinsing
5. Disinfection
6. Final rinsing
7. Inspection
8. Packaging
Operating mode
- Rinse manually and
abundantly the
accessories with current
water
- clean manually the
accessories with an
adequate brush,
preliminary soaked in a
clean pre-disinfectant
solution
See point # 2
- Immerse the
accessories in a
disinfectant solution
(bactericidal, virucidal,
fungicidal, tuberculocidal
and aldehyde free)
according to the
manufacturer
recommendations
- See point # 2
- After rinsing, the
accessories have to be
dried.
- Inspect devices and sort
out those with defects.
- Pack the devices in
"Sterilization pouches".
Warning
- the hook mechanism has to be
activated during the cleaning
process (press several times the
push button)
- no visible impurities should be
observed on the accessories
- Follow instructions and observe
concentrations and time given by
the manufacturer
- Dirty instruments must be
cleaned and disinfected again.
- Check the validity period of the
pouch given by the manufacturer
to determine the shelf life.
- Use packaging which are
resistant up to a temperature of
141°C (286°F) and in accordance
with EN ISO 11607.
20 / 112