As a safety precaution in order to avoid over-instrumentation, it is recommended
to proceed as follows: place the file onto an endodontic ruler at the point where
the ProPex
®
measured file length.
Please also respect the following precautions:
-
For your own safety, be aware of wearing personal protective equipment
(gloves, glasses, mask).
-
If the bar graph makes sudden large movements in the coronal part of the
canal, slowly continue advancing the file toward the apex until the signal
returns to normal.
-
This ProPex
other apparatus or system. It must not be used as an integral component of
any other apparatus or system. Using replacement parts or accessories not
supplied by the original manufacturer or vender could adversely affect the
EMC performance of the ProPex
-
The device must be used with the manufacturer's original accessories only.
-
Unplug the device before replacing the battery.
-
Never use batteries that are leaky, deformed, discolored or otherwise
abnormal.
-
In case of battery leakage, carefully dry the battery terminals and remove all of
the leaked liquid. Then replace the battery with a new one.
-
Dispose of old batteries according to local codes and regulations.
-
Accessories including lip clips, hooks or forks should be clean and without
residue of chemical disinfectants or other medicinal solutions such as sodium
hypochlorite or formalin.
-
Do not expose ProPex
-
®
ProPex
II must be stored in normal temperature (< 60°C) and humidity
conditions.
5. Adverse Reactions
If the apex locator provides incorrect reading and there is no radiographic data (see
"Important notice" in the "Precautions" section in chapter 4), the following adverse
reactions may occur:
Incomplete root canal treatment,
Apex perforation.
II indicates 'APEX'. Subtract a minimum of 0.5 mm from the
®
II unit must not be connected to or used in combination with any
®
II to any liquid.
®
II.
9 / 112