EN
Set-up and Start-up
Locking buttons
22
1.
Remove the vapor tube from the packaging.
3
4
Figure 4-7: Install the vapor tube
2.
Remove the threaded clamping screw (1) on the drive unit; remove the tension
spring (2) and PTFE 26 vacuum seal (3).
Figure 4-8: Locking button
3.
Continue to press the locking button (Figure 4-8 (1)).
4.
Remove the screw coupling connection of the vapor tube (Figure 4-7 (6) from the
drive unit.
5.
Pull out the clamping sleeve (Figure 4-7 (5).
6.
Push the vapor tube coupling clamp (Figure 4-7 (6)) and clamping sleeve (Figure 4-
7 (5)) with the bead first to the vapor tube (Figure 4-7 (7)). Verify that the bead audi-
bly snaps into the groove of the vapor tube (Figure 4-7 (7)).
7.
Push the vapor tube (Figure 4-7 (7) into the drive unit with the pre-assembled parts.
8.
Continue to press the locking button (Figure 4-8 (1)).
9.
Manually tighten the vapor tube coupling clamp (Figure 4-7 (6)) with the threaded
connection of the motor drive (Figure 4-7 (4)) till a tight connection is made.
1
2
5
6
Hei-VAP Value, Hei-VAP Advantage
7