How to Test Glucose Control
Use ONLY TRUEtest Glucose Control Solution
with the TRUE2go.
1. Allow Control, vial of Strips and Meter to adjust to
68-77ºF (room temperature). Write date first opened
on Control bottle label and Strip vial label.
Note: Running a Control Test at temperatures
outside the range listed above may cause
Control to read as a blood test.
2. Check dates on Control label and Strip
vial label. Do not use Control or Strips if
Expiration Dates have passed (Control
- 3 months after opening or date next
to EXP on label, Strips - 4 months after
opening or date next to EXP on label).
3. Swirl or invert bottle gently to
mix Control. DO NOT SHAKE!
4. Open vial by pushing up under the
Test Port of the Meter. Remove one Strip
from vial. Close vial immediately by firmly
pressing down on the top of the Meter.
Note: Use Strip quickly after removal from
vial.
5. Insert Strip into Test Port. Meter
turns on.
Note: If Strip has been out of the vial too
long before testing, an error message
appears upon insertion of the Strip
into the Meter. Release and discard
old Strip. Use new Strip for testing.
17
May 30, 2012
Press Down
To Close
Lift Up
To Open
Contacts
face up