Transmisor de presión inteligente STR800
H2
Fixed
ref. leg
NOTE: Lower flange seal should not be mounted over 22 feet below or above
the transmitter for silicone fill fluid (11 feet for CTFE fill fluid) with tank at one
atmosphere. The combination of tank vacuum and high pressure capillary
head effect should not exceed 9 psi vacuum (300 mmHg absolute).
Consult Honeywell for installation of STR13D.
Figura 6: el transmisor STR800 con sello de diafragma remoto montado en un tanque
LP Side
Maximum level
Minimum level
.
HP Side