Fault
Cause
Pump without function
Pump is blocked or
defective
Boil-dry protection is
triggered
The hold button does
Hold button is defective
not react
Water temperature of
65 °C not reached yet
6-min timer without
6-min timer is blocked
return function, pump
running permanently
6-min timer is defective
66
Remedy
Examine the pump,
replacing if necessary
Activate the boil-dry
protection function
(see instructions for
activation)
Replace the hold button
Wait until temperature
level of 65 °C is reached
Eliminate blockage,
re-centering the switch
if necessary
Examine the timer,
replacing if necessary
Only Labormat TH (
D3603N / D3603NUSA)
Fault
Cause
Pump is leaky
Water level is too high
in the tank
Four-hole seal
(between pump and
basin) is defective
Only Labormat SD (
D3601 / D3601USA)
Fault
Cause
24-h timer without /
24-h timer is defective
with limited return
function
24-h timer is defective
Remedy
Reduce water to max.
level, observe marking,
max. 15 litres
Dismantle pump and
replace four-hole seal
Remedy
Compare the power
source with the voltage
supply indicated,
adapting if necessary
230 V~/ 50Hz
220 V~/ 60Hz
Eliminate blockage,
re-centering the switch
if necessary