4. REPROCESSING

The AIRVO 2 must be cleaned and disinfected between patients according to the instructions in the
Disinfection Kit Manual (900PT600).
This should take place as soon as possible after use. The unit utilizes warmed water and can pose a risk of
bacterial colonization and patient infection if cleaning, disinfection and replacement procedures are not
followed.
Standard aseptic techniques to minimize contamination should be followed when handling the unit
and accessories. This includes proper hand-washing, avoiding hand contact with connection ports, safe
disposal of the used consumables and suitable storage of the unit after cleaning and disinfection.

SCHEDULE FOR CHANGING ACCESSORIES

The accessories for the unit must be changed frequently to avoid the risk of infection. Parts should be
replaced immediately if they are damaged or discolored; otherwise they must be replaced within the
periods shown in the following table.
Maximum
period of use
Patient interfaces excluding Optiflow™+
OPT316
OPT318
1 week
OPT842
(single-patient
OPT844
use)
OPT846
OPT870
RT013
Optiflow™+ patient interfaces
OPT942
OPT944
OPT946
OPT970
2 weeks
OPT980
(single-patient
use)
All tube & chamber kits
900PT551 /
900PT561
900PT501
900PT531
900PT913
3 months
or
1000 hours

FILTER REPLACEMENT

After the AIRVO 2 has been switched on for 1000 hours, a prompt will appear at the start of the next
disinfection cycle indicating that an air filter change is due. Follow the steps below if filter change is due:
Air filter change
due
Replace now?
Now
Later

SERVICING

This device contains no internal serviceable parts.
AIRVO 2 humidifiers do NOT require routine servicing or calibration.
Refer to the AIRVO 2 Technical Manual for a list of external spare parts.
Part number and description
Nasal Cannula - Infant
Nasal Cannula - Pediatric
Optiflow™ Nasal Cannula - Small
Optiflow™ Nasal Cannula - Medium
Optiflow™ Nasal Cannula - Large
Tracheostomy Interface
Mask Interface Adapter - 22mm
Optiflow™+ Nasal Cannula - Small
Optiflow™+ Nasal Cannula - Medium
Optiflow™+ Nasal Cannula - Large
Optiflow™+ Tracheostomy Interface
Optiflow™+ Mask Interface Adapter
AirSpiral™ Heated breathing tube, MR290 auto-fill chamber and adapter
Heated breathing tube, MR290 auto-fill chamber and adapter
Junior heated breathing tube, MR290 auto-fill chamber and adapter
(for use with OPT316/318 only)
Air filter
(or more often if significantly discolored)
1.
Take the filter holder from the back of the unit and remove the filter.
2.
Replace the old filter with a new filter (900PT913).
3.
Reattach the filter holder to the unit (clip the bottom of the filter
holder in first, then rotate it upwards until the top clips into place).
4.
Press the Mode button to move on to the "Replace now" screen.
5.
Press the Up button to select "Now".
6.
Press the Mode button to confirm.
The hours counter will be reset to zero.
If you choose the "Later" option, the prompt will continue to appear at
the start of subsequent disinfection cycles.
A – 12