Guidance and manufacture's declaration-electromagnetic immunity for all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration-electromagnetic immunity
The MQ3000 Pulse Oximeter is intended for use in the electromagnetic environment specified specified below.
The the user of MQ3000 Pulse Oximeter should assure that it is used in such an environment.
Immunity test
Electrostatic dis-
charge (ESD) IEC
61000-4-2
P o w e r f r e q u e n c y
( 5 0 H z ) magnetic
field IEC 61000-4-8
IEC60601 test level
±6KV contact
±8KV air
3A/m
Compliance level
Electromagnetic environment-guidance
Floors should be wood, concrete or ceramic tile.
±6KV contact
If floor are covered with synthetic material, the
±8KV air
relative humidity should be at least 30%.
3A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment