Safety cut out
If the internal heater temperature exceeds normal operating temperature, a safety cut out will operate and the
display will flash OTC. Switch off the unit and contact GRANT or your local GRANT representative for
support. To avoid this fault, always use the circulation tray provided and ensure the bath does not run
dry.
Maintenance and Service
Cleaning
Clean the bath with a damp cloth, using only water. Do not use chemical cleaning agents.
Scale on immersed parts can be removed using chemical de-scaling products designed for use on kitchen
equipment which has metal parts. Warning - De-scaling products may be toxic - follow the manufacturer's
instructions.
Conditions of use
The water baths are for indoor use only. Check that the environmental conditions are within the following
limits:
Temperature
Maximum relative humidity
Altitude
Electrical details
Mains supply
220-240Vac @ 50/60 Hz or 110-120Vac @ 50/60Hz
Pollution degree: 2
Installation Category: II
Note: Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage
Model
SVP 5
SVP 12
SVP 26
Replacement of fuses
Disconnect unit from the power supply before changing fuses. Replace fuses only by the same type and rat-
ing, see above table. They must be 1.25 inch, High Breaking Capacity (HBC) type.
Guarantee
When used according to these operating instructions, these baths are guaranteed for 1 year against faulty
materials or workmanship.
Service
There are no user serviceable parts inside the unit.
If potentially hazardous liquid is spilt onto the equipment, disconnect it from the power supply and have it
checked by a competent person. It is the user's responsibility to carry out appropriate decontamination if
hazardous material is spilt on the equipment.
For service or repair please contact the Grant Service Department in the UK.
Service Address:
o
5 to 40
C
80 % r.h. in room temperatures up to 31°C decreasing linearly to 50 % r.h. at 40 °C
Up to 6,500 feet (2,000 m) above sea level
Capacity
Current Consumption
Litres
220-240V
5L
1.5A
12L
2.9A
26L
5.5A
Service Department,
Grant Instruments (Cambridge) Ltd
Shepreth, Cambridgeshire,
SG8 6GB, England.
Tel:
+44 (0) 1763 260 811
Fax:
+44 (0) 1763 262 410
E-mail:
labservice@grant.co.uk
110-120V
220-240V
2.7A
5.5A
8.6A
Fuse rating
110-120V
5AF
5AF
10AF
10AF
15AF
15AF